Bead Cleanup and dsDNA Quantification
Bead Cleanup
Dual selection to remove both high (gDNA) and low (primer/dimer) MW fragments
- Mix 100ul beads per 200ul PCR reaction by pipetting (0.5:1 ratio).
- Incubate 15 min at RT with mixing.
- Incubate on magnet 10 min.
- Remove 275ul solution to new tube while tube is still attached to magnet (large MW fragments remain on beads and are discarded).
- Add 55ul beads per 275ul eluate (0.8:1 ratio, accounting for transferred solute).
- Incubate 15 min at RT with mixing.
- Incubate on magnet 10 min.
- Remove and discard supernatant containing primer/dimer fragments. Keep the beads!
- Wash twice with 200ul fresh 70% ethanol while tube is still attached to magnet (do not disturb the beads):
Add 200ul 70% ethanol. Let sit 30 sec RT on magnet. Carefully remove ethanol wash. Repeat.
*If using a large magnet, tilt on its side so ethanol covers beads and rock gently. - After removal of second wash, air dry beads 1 min.
- Resuspend beads well in 35ul H2O.
- Incubate 2 min at RT; place on magnet.
- Run 2ul on a 1-2% TAE agarose gel to verify removal of unwanted fragments.
Measuring dsDNA Concentration with Qubit
- Equilibrate all High Sensitivity dsDNA Qubit reagents to RT.
- Set up Qubit assay tubes: sample number +2 for standards.
- Prepare working solution in a plastic tube (200ul per sample + 200ul extra - no glass!). 1:200 reagent:buffer
- Aliquot working solution to tubes; add standard or sample. Vortex 2-3 sec.
200ul total per tube:
For standard: 190ul working solution + 10ul standard
For sample: 190-199ul working solution + 1-20ul sample - Incubate at RT at least 2 min.
- Read on Qubit.